By default device is not set for continuous sampling - configure as required but be aware readings may not update like fixed devices on continuous operation mode. Users have a choice of standard data interface options including RS-485 Modbus, or 4-20 mA. The CI-3100 RS requires both an external vacuum pump and DC power. Use front screen to configure IP address The CI-3100 RS (Remote Sensor) Series is a 2-channel particle counter used in continuous monitoring applications in a variety of industries generally Grade A or B (ISO 5) clean zones. First the devices must have IP address set Connect to network and add to RMS as normal (example Modbus map attached) > set inst auto 1 (sets device yo auto start) > set sample 60, 60 (sets sample time to 60seconds. > set queue 1 (limits data queue to 1 - always the live value) > set mode 2 (sets device to modbus mode) It is the first particle counter with a built-in vacuum pump offered with Power-Over-Ethernet (POE+). BioPharmaceutical grade, a 2-channel (0.5 m and 5.0 m) particle counter with a 10,080 sample memory buffer, integrated light ring, and enhanced diagnostics. Use "set ip", "set mask" and "set gateway" commands to set IP settings (as per manual) CI-3100 TRIDENT Non-Viable Particle Counter. Use putty as per manual to connect to ISOAIR Connect cable to USB port on PMS device (using supplied cable) and to PC using generic Serial to USB convertor cable First the devices must have IP address reset via Serial Cable.